Overview
Endress+Hauser Proline Promag P300 Electromagnetic Flowmeter is a high-temperature flowmeter for process applications with a compact, easily accessible transmitter.
Specs at a glance
* Max. measurement error
Volume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s), Flat Spec
* Measuring range :4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min)
* Medium temperature range
Liner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PFA high-temperature: –20 to +180 °C (–4 to +356 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
* Max. process pressure :PN 40, Class 300, 20K
* Wetted materials
Liner: PFA; PTFE
Electrodes: 1.4435 (F316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium
Benefits
* Diverse applications – wide variety of wetted materials
* Energy-saving flow measurement – no pressure loss due to cross-section constriction
* Maintenance-free – no moving parts
* Full access to process and diagnostic information – numerous, freely combinable I/Os and Ethernet
* Reduced complexity and variety – freely configurable I/O functionality
* Integrated verification – Heartbeat Technology
Application
Promag P is dedicated to chemical and process applications with corrosive liquids and highest medium temperatures. With its compact transmitter Promag P 300 offers a high flexibility in terms of operation and system integration: access from one side,remote display and improved connectivity options. Heartbeat Technology enables compliance and process safety at all times.
* The bidirectional measuring principle is virtually independent of pressure, density, temperature and viscosity
* Dedicated to chemical and process applications with corrosive liquids and high medium temperatures
Specification
Model Number | Proline Promag P 300 |
Measuring principle | Electromagnetic |
Product headline | High-temperature flowmeter for process applications with a compact, easily accessible transmitter. Dedicated to chemical and process applications with corrosive liquids and high medium temperatures. |
Sensor features | Diverse applications – wide variety of wetted materials. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts. Nominal diameter: max. DN 600 (24"). All common Ex approvals. |
Transmitter features | Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology. Liner made of PTFE or PFA. Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. |
Nominal diameter range | DN 15 to 600 (½ to 24") |
Wetted materials | Liner: PFA; PTFE Electrodes: 1.4435 (F316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum; Titanium |
Measured variables | Volume flow, conductivity, mass flow |
Max. measurement error | Volume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s) Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s), Flat Spec |
Measuring range | 4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min) |
Max. process pressure | PN 40, Class 300, 20K |
Medium temperature range | Liner material PFA: –20 to +150 °C (–4 to +302 °F) Liner material PFA high-temperature: –20 to +180 °C (–4 to +356 °F) Liner material PTFE: –40 to +130 °C (–40 to +266 °F) |
Ambient temperature range | Flange material carbon steel: –10 to +60 °C (+14 to +140 °F) Flange material stainless steel: –40 to +60 °C (–40 to +140 °F) |
Sensor housing material | DN 15 to 300 (½ to 12""): AlSi10Mg, coated DN 350 to 600 (14 to 24""): Carbon steel with protective varnish |
Transmitter housing material | AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; stainless steel for hygenic transmitter design |
Degree of protection | Standard: IP66/67, Type 4X enclosure |
Display/Operation | 4-line backlit display with touch control (operation from outside) Configuration via local display and operating tools possible Remote display available |
Outputs | 3 outputs: 4-20 mA HART (active/passive) 4-20 mA WirelessHART 4-20 mA (active/passive) Pulse/frequency/switch output (active/passive) Double pulse output (active/passive) Relay output |
Inputs | Status input 4-20 mA input |
Digital communication | HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA |
Power supply | DC 24 V AC 100 to 230 V AC 100 to 230 V / DC 24 V (non-hazardous area) |
Hazardous area approvals | ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC |
Product safety | CE, C-tick, EAC marking |
Functional safety | Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511 |
Metrological approvals and certificates | Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025) Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) |
Marine approvals and certificates | LR approval, DNV approval, ABS approval, BV approval, CCS approval |
Pressure approvals and certificates | PED, CRN |
Material certificates | 3.1 material |
Hygienic approvals and certificates | ACS, NSF 61, WRAS |
We normally ship by DHL, UPS, FEDEX or TNT,