Overview
Economical Ultrasonic flowmeter with integrated temperature measurement.
Specs at a glance
* Max. measurement error
Volume flow:
- ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)
* Measuring range
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min)
* Medium temperature range
0 to 150° (17.8 to 302 °F)
* Max. process pressure
PN 25 / ASME Cl. 150
* Wetted materials
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105
Volume flow:
- ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)
* Measuring range
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min)
* Medium temperature range
0 to 150° (17.8 to 302 °F)
* Max. process pressure
PN 25 / ASME Cl. 150
* Wetted materials
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105
Application
Measuring of demineralized water in utilities, for example in boiler condensate return lines, requires reliable devices. Prosonic Flow E 100 provides economical and accurate bidirectional flow measurement across all industries. With its integrated temperature measurement, Prosonic Flow E 100 proves to be a multivariable flowmeter for considerable additional value in your plant. Heartbeat Technology enables measurement reliability and extension of recalibration cycles.
* The measuring principle is independent of pressure, density, temperature and conductivity
* Bidirectional measuring of demineralized water applications for utilities, e.g. boiler condensate return lines. Device properties:
* Accuracy up to ±0.5 % (flow) or acc. to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temp.)
* Process temperatures up to 150 °C (302 °F)
* Entire meter housing made of stainless steel
* 4-20 mA HART, pulse/frequency output
* Local display for reading and monitoring available
* Robust transmitter housing
* The measuring principle is independent of pressure, density, temperature and conductivity
* Bidirectional measuring of demineralized water applications for utilities, e.g. boiler condensate return lines. Device properties:
* Accuracy up to ±0.5 % (flow) or acc. to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temp.)
* Process temperatures up to 150 °C (302 °F)
* Entire meter housing made of stainless steel
* 4-20 mA HART, pulse/frequency output
* Local display for reading and monitoring available
* Robust transmitter housing
Benefits
* Long-term stability – reliable, robust sensor
* Reducing further measuring point – multivariable device
* Dependable flow measurement – high turndown (200:1)
* Time-saving local operation without additional software and hardware – integrated web server
* Extended calibration intervals – integrated device verification due to Heartbeat Technology
* Easy commissioning – brief parameter explanations
* Reducing further measuring point – multivariable device
* Dependable flow measurement – high turndown (200:1)
* Time-saving local operation without additional software and hardware – integrated web server
* Extended calibration intervals – integrated device verification due to Heartbeat Technology
* Easy commissioning – brief parameter explanations
Specification
item
|
value
|
Warranty
|
1 Year
|
Customized support
|
OEM
|
Place of Origin
|
Germany
|
Brand Name
|
Endress+Hauser
|
Model Number
|
Prosonic Flow E100
|
Type
|
Ultrasonic flow meters
|
Product Headline
|
Economical Ultrasonic flowmeter with integrated temperature measurement.
Bidirectional measuring of demineralized water applications for Utilities, e.g. boiler condensate return lines. |
Sensor Features
|
Long-term stability – reliable, robust sensor. Reducing further measuring point – multivariable device. Dependable flow
measurement – high turndown (200:1). Measurement accuracy up to ±0.5 % (flow) or according to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temperature). Process temperatures up to 150 °C (302 °F). Entire meter housing made of stainless steel. |
Transmitter Features
|
Time-saving local operation without additional software and hardware – integrated web server. Extended calibration intervals –
integrated device verification due to Heartbeat Technology. Easy commissioning – brief parameter explanations. 4-20 mA HART, pulse/frequency output. Local display for reading and monitoring available. Robust transmitter housing. |
Nominal Diameter Range
|
DN 50 to 150 (2"to 6")
|
Wetted Materials
|
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105 |
Measured Variables
|
Volume flow, calculated mass flow, flow velocity, sound velocity, temperature
|
Max. Measurement Error
|
Volume flow:
- ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s) - ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s) |
Measuring Range
|
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min) |
Max. Process Pressure
|
PN 25 / ASME Cl. 150
|
Medium Temperature Range
|
0 to 150° (17.8 to 302 °F)
|
Ambient Temperature Range
|
-20 to 60 °C (-4 to 140 °F)
|
Sensor Housing Material
|
Stainless steel
|
Transmitter Housing Material
|
Compact:
AlSi10Mg, coated Stainless steel, 1.4301 |
Degree Of Protection
|
Standard: IP66/67, Type 4X enclosure
With opened housing: IP20, Type 1 enclosure |
Display/Operation
|
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible |
Outputs
|
4‐20 mA HART (active)
Pulse/frequency/switch output (passive) |
Inputs
|
None
|
Power Supply
|
DC 19.2 to 28.8 V
|
Hazardous Area Approvals
|
None
|
Product Safety
|
CE, C-Tick, EAC marking
|
Metrological Approvals And Certificates
|
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) Calibration according to EN 1434 Class 2 |
Pressure Approvals And Certificates
|
PED
|
Material Certificates
|
3.1 material
|
please contact email for more about the shipping cost
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.