Self-calibrating hygienic temperature transmitter lowers risk and drives automation in regulated processes.
Specs at a glance
-
Response time t50 = 2.5 s t90 = 5.4 s
-
Max. process pressure (static) at 20 °C: 40 bar (580 psi)
-
Operating temperature rangePt100:-40 °C to 160 °C (-40 °F to 320 °F),
optional up to 190 °C (374 °F) reference point for automated calibration -
Max. immersion length on request up to 900.00 mm (35.4'')
Benefits
* True RTD sensor self-calibration making use of the Curie effect; Built-in, long-term stable fix point reference with complete traceability of calibration chain to ITS-90
* Full instrument diagnostics, monitoring and verification with Heartbeat Technology
* Automated process documentation, built-in memory for 350 calibration events; 24/7 access to printable, audit proof calibration certificates, on-site or through cloud-based system integration
* Built-in 4 to 20 mA loop check function saves time and increases process safety
* Highest temperature measurement precision: individual transmitter-sensor matching from factory for improved RTD accuracy
-
Application
production downtime while increasing product safety and process efficiency in food, beverage and life sciences applications. The instrument complies with FDA and GMP regulations and integrates with the Netilion cloud-based Industry 4.0 ecosystem.
* RTD self-calibration in processes such as SIP, UHT, edible oil refining, industrial deep fat frying
* Measuring range: -40 to 190 °C (-40 to 374 °F)
* Protection class: IP65/67 or IP69* Signal output: 4 to 20 mA, HART®
* Optional: Secure Bluetooth® and/or WirelessHART® (requires FieldPort SWA50)
* EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE, CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
Specification
Warranty |
1 Year |
Customized support |
OEM, ODM, OBM |
Place of Origin |
Germany |
Brand Name |
ENDRESS+HAUSER |
Model Number |
TM371 |
Theory |
Temperature Transmitter |
Product name |
Sanitary RTD - iTHERM TrustSens TM371 |
Measuring range |
-40 to 190 °C (-40 to 374 °F) |
Measuring principle |
Resistance Temperature Detector |
Characteristic / Application |
self-calibrating metric style compact temperature probe hygienic design life sciences, food & beverage |
Outer diameter protection tube / Insert |
6.0 mm (0.24'') 9.0 mm (0.35'') 12.7 mm (0.5") |
Max. immersion length on request |
up to 900.00 mm (35.4'') |
Material protection tube/ thermowell |
1.4435 Delta Ferrite < 1% 316L |
Operating temperature range |
Pt100: -40 °C to 160 °C (-40 °F to 320 °F), optional up to 190 °C (374 °F) |
Max. process pressure (static) |
at 20 °C: 40 bar (580 psi) |
Response time |
t50 = 2.5 s t90 = 5.4 s |