Endress+Hauser Ultrasonic measurement Time-of-Flight Prosonic FDU90

: In stock

: Level Measurement

: Endress+Hauser

Regular price Sale

Overview Ultrasonic sensor for level and flow measurement for connection to FMU9x (measuring range up to 3m). Specs at a glance * Accuracy+/- 2mm + 0.17% of measured...


Overview

Ultrasonic sensor for level and flow measurement for connection to FMU9x (measuring range up to 3m).
Specs at a glance
* Accuracy
+/- 2mm + 0.17% of measured distance
* Process temperature
-40 °C ... 80 °C
(-40 °F ... 176 °F)
* Process pressure / max. overpressure limit
0.7 bar ... 4 bar abs
(10 psi ... 58 psi)
* Max. measurement distance
Liquids: 3 m (9.8 ft),
Solids: 1.2 m (3.9 ft)
* Main wetted parts
PVDF (IP68 / NEMA6P)

Application

The FDU90 ultrasonic sensor for continuous, non-contact and maintenance-free level measurement of fluids, pastes, sludges and powdery to coarse bulk materials. But also for flow measurement in open channels and measuring weirs. The measurement is unaffected by dielectric constant, density or humidity and also unaffected by build-up due to the self-cleaning effect of sensors.
Suited for explosion hazardous areas. Maximum measuring range in liquids 3m (9.8ft), solids 1.2m (3.9ft).
Sensor for continuous, non-contact level measurement of fluids and bulk materials in silos, on belts, stockpiles and in crushers.
Also flow measurement in open channels and at weirs. Typically used in measurements in low structural heights.
* Process connection: Thread
* Temperature: -40 to +80°C (-40 to +176°F)
* Pressure: +0.7 to +4bar (+10 to +58psi)
* Maximum measuring range: Liquids 3m (9.8ft), solids 1.2m (3.9ft)
* Blocking distance: 0.07m (0.23ft)
* International explosion protection certificates

Benefits

* Integrated temperature sensor for Time-of-Flight correction. Accurate measurements are possible, even if temperature changes are present
* Hermetically welded PVDF sensor for highest chemical resistance
* Suited for rough ambient conditions thanks to separate installation from the transmitter (up to 300m)
* Reduced build-up formation because of the self-cleaning effect
* Integrated automatic sensor detection for transmitters FMU90 /FMU95 for simple commissioning
* Weather resistant and flood-proof (IP 68)
* Integrated heating against a build-up of ice at the sensor (optional) ensures reliable measurement

Specification

item
value
Grade
Industrial
Warranty
1 Year
Customized support
OEM
Place of Origin
Germany
Brand Name
Endress+Hauser
Model Number
Prosonic FDU90
Material
PVDF
Mode
Ultrasonic measurement Time-of-Flight
Measuring Principle
Ultrasonic
Characteristic / Application
Separated version with field housing or top hat rail housing for control cabinet instrumentation, 300m in-between sensor and
transmitter
Specialities
Floating protection tube
Supply / Communication
4-wire (HART, Profibus DP)
Accuracy
+/- 2mm + 0.17% of measured distance
Ambient Temperature
-40 °C ... 80 °C (-40 °F ... 176 °F)
Process Temperature
-40 °C ... 80 °C (-40 °F ... 176 °F)
Process Pressure / Max. Overpressure Limit
0.7 bar ... 4 bar abs (10 psi ... 58 psi)
Main Wetted Parts
PVDF (IP68 / NEMA6P)
Process Connection
Front side thread:
G / NPT 1 1/2"
Rear side thread:
G / NPT 1"
Ceiling mounting
Blocking Distance
0.07 m (0.23 ft)
Max. Measurement Distance
Liquids: 3 m (9.8 ft),
Solids: 1.2 m (3.9 ft)
Communication
4 ... 20 mA HART Profibus DP
Certificates / Approvals
ATEX, FM, CSA, IEC Ex, INMETRO, NEPSI, EAC Ex
Options
Second 4...20mA output
Components
Transmitter:
FMU90, FMU95
Application Limits
Foam / high turbulence possible:
FDU92
Flange-flush assembly:
FDU91F
For tank farms scanner:
FMU95


please contact email for more about the shipping cost

We normally ship by DHL, UPS, FEDEX or TNT,

some country could not be ship by above method, then we would use EMS or Country Logistics line.