Endress+Hauser Proline t-mass T150 thermal mass flowmeter is a flowmeter for reliable and easy monitoring of liquids.
Specs at a glance
* Max. measurement error
Flow: ±5 % o.f.s.
* Measuring range
226 to 14 100 000 l/h (60 to 3 730 000 gal/h)
(under reference conditions)
* Medium temperature range
–20 to +100 °C (–4 to +212 °F)
* Max. process pressure
PN 40
* Wetted materials
Transducer: 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022)
Insertion tube: 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022)
Connection:
‐ Compression fitting: 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022)
‐ Threadolet: 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022)
‐ Clamping ferrule: PEEK 450G; 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022)
‐ Tri‐Clamp; DN40 DIN 11851, DN50 DIN 11851; DN40 DIN 11864‐1A, DN50 DIN 11864‐1A: 1.4404 (316L)
Application
The t-mass T 150 is the first thermal mass device from Endress+Hauser for measuring liquids. t-mass T 150 is designed chiefly for water applications. As it measures independently of the electrical conductivity of a fluid and can be used in a variety of water-based and non-water-based liquids for the purpose of monitoring and trending. Customer-specific settings are saved on the display and can be transferred from one device to another by means of the display.
* Measuring principle is characterized by a high operable flow range and direct mass flow measurement
* Dedicated to the monitoring of conductive and non-conductive liquids.
Device properties
* Insertion version for nominal diameter DN 40 to 1000 (1 1/2 to 40")
* Sensor in standard or hygienic version
* SIP cleaning possible up to 130 °C (266 °F)
* Device in compact version with DC 24 V power supply
* 4-20 mA HART, pulse/frequency/switch output
* Compact and robust transmitter
Benefits
* Dedicated to the monitoring of conductive and non-conductive liquids
* High process safety – high repeatability and linearity due to integrated temperature compensation
* Cost-effective measurement – easy installation, negligible pressure loss and maintenance-free
* Reliable flow trending – multivariable measurement
* Fast and efficient commissioning – guided operating menus
* High plant availability – self-diagnostics and error monitoring
* Automatic recovery of data for servicing
Specification
Measuring Principle | Thermal |
Product Headline | The flowmeter for reliable and easy monitoring of liquids. Dedicated to the monitoring of conductive and non‐conductive liquids. |
Sensor Features |
High process safety – high repeatability and linearity due to integrated temperature compensation. Cost‐effective measurement – easy installation, negligible pressure loss and maintenance‐free. Reliable flow trending – multivariable measurement. Insertion version for nominal diameter DN 40 to 1000 (1½ to 40"). Sensor in standard or hygienic version. |
Transmitter Features | Fast and efficient commissioning – guided operating menus. High plant availability – self‐diagnostics and error monitoring. Automatic recovery of data for servicing. Device in compact version with DC 24 V power supply. 4‐20 mA HART, pulse/frequency/switch output. |
Nominal Diameter Range | DN 40 to 1000 (1½ to 40") |
Wetted Materials | Transducer: 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022) Insertion tube: 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022) Connection: ‐ Compression fitting: 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022) ‐ Threadolet: 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022) ‐ Clamping ferrule: PEEK 450G; 1.4404 (316L); Alloy C22, 2.4602 (UNS N06022) ‐ Tri‐Clamp; DN40 DIN 11851, DN50 DIN 11851; DN40 DIN 11864‐1A, DN50 DIN 11864‐1A: 1.4404 (316L) |
Measured Variables | Mass flow, temperature, volume flow |
Max. Measurement Error | Flow: ±5 % o.f.s. |
Measuring Range | 226 to 14 100 000 l/h (60 to 3 730 000 gal/h) (under reference conditions) |
Max. Process Pressure | PN 40 |
Medium Temperature Range | –20 to +100 °C (–4 to +212 °F) |
Ambient Temperature Range | –40 to +60 °C (–40 to +140 °F) |
Transmitter Housing Material | AlSi10Mg, coated |
Degree Of Protection | IP66/67, type 4X enclosure |
Display/Operation | 4‐line display with push buttons Configuration via local display and operating tools possible |
Outputs | 4‐20 mA HART (active) Pulse/frequency/switch output (passive) |
Inputs | Status input |
Digital Communication | HART |
Power Supply | DC 18 to 30 V |
Hazardous Area Approvals | ATEX, IECEx, cCSAus |
Other Approvals And Certificates | 3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR Hygienic approvals: EHEDG, 3-A |
Metrological Approvals And Certificates | Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR |
Pressure Approvals And Certificates | CRN |
Material Certificates | 3.1 material |
Hygienic Approvals And Certificates | Sanitary approval: 3-A, EHEDG |
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.