Overview
100% compliance, 0% effort Hygienic compact RTD thermometer, US style.
Specs at a glance
Response time
t50 = 2,5 s
t90 =9,5s
Max. process pressure (static)
at 20 °C: 40 bar (580 psi)
Operating temperature range
PT 100:
-40 °C …160 °C (-40 °F …320 °F),
optional up to 190 °C (374 °F)
Max. immersion length on request
up to 28'' (711 mm)
others on request
Response time
t50 = 2,5 s
t90 =9,5s
Max. process pressure (static)
at 20 °C: 40 bar (580 psi)
Operating temperature range
PT 100:
-40 °C …160 °C (-40 °F …320 °F),
optional up to 190 °C (374 °F)
Max. immersion length on request
up to 28'' (711 mm)
others on request
Field of application
The award-winning smart temperature probe with integrated HART® transmitter and Heartbeat Technology features the world’s first
RTD sensor unit capable of fully automated in-situ self-calibration. It effectively eliminates the risk of undetected non-conformities, reduces production downtime and increases product safety and process efficiency in applications of the food & beverage and life sciences industries. The device is fully compliant to FDA 21 CFR Part 11 regulations and GMP rules.
* Specially designed for use in hygienic and aseptic applications in the food & beverage and life sciences industries
* Measuring range: –40 to 190 °C (–40 to 374 °F)
* Pressure range up to 50 bar (725 psi)
* Protection class: IP65/67 or IP69
* Communication: Analog output 4 to 20 mA, HART® protocol
* Measuring range: –40 to 190 °C (–40 to 374 °F)
* Pressure range up to 50 bar (725 psi)
* Protection class: IP65/67 or IP69
* Communication: Analog output 4 to 20 mA, HART® protocol
Benefits
* Risk and cost reduction thanks to self-calibration and Heartbeat Technology
* No production downtime due to an automated and fully traceable inline self-calibration
* Automatized documentation, memory for 350 calibration points.
Printable calibration certificate - audit proof.
* Automated 4 to 20 mA loop check for saving time and increasing safety in your process.
* Elimination of non-conformities or undetected failures
* International certifications and approvals:
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
* Highest accuracy of measuring point through sensor-transmitter matching
* Automatized documentation, memory for 350 calibration points.
Printable calibration certificate - audit proof.
* Automated 4 to 20 mA loop check for saving time and increasing safety in your process.
* Elimination of non-conformities or undetected failures
* International certifications and approvals:
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
* Highest accuracy of measuring point through sensor-transmitter matching
Specification
Brand Name
|
ENDRESS+HAUSER
|
Model Number
|
TM372
|
Theory
|
Temperature Transmitter
|
Measuring principle
|
Resistance Temperature Detector
|
Measuring range
|
–40 to 190 °C (–40 to 374 °F)
|
Pressure range
|
at 20 °C: 40 bar (580 psi)
|
Protection class
|
IP65/67 or IP69
|
Response time
|
t50 = 2,5 s
t90 =9,5s |
Operating temperature range
|
PT 100:
-40 °C …160 °C (-40 °F …320 °F), optional up to 190 °C (374 °F) |
Characteristic Application
|
self-calibrating
imperial style compact temperature probe hygienic design hygienic process connection |
Thermowell protection tube
|
without
incl. thermowell incl. Tee-/elbow thermowell |
Insert probe
|
pipe version, isolated wires, not flexible
|
Outer diameter protection tube Insert
|
1/4'' (6,35 mm)
3/8'' (9,53 mm) 1/2" (12,7 mm) |
Max. immersion length on request
|
up to 28'' (711 mm)
|
Material protection tube/ thermowell
|
316L
|
Process connection
|
Weld-in adapter
Tri-clamp Clamp-connections according to ISO2852 G3/4" Liquiphant G1" Liquiphant Varivent T- and elbow thermowell |
Tip shape
|
straight
reduced |
Integration head transmitter
|
no (4…20mA Signal)
|
Ex - approvals
|
CSA C/US IS, I/1+2/A-F
|
International certifications and approvals:
|
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
|
please contact email for more about the shipping cost
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.