Overview
For continuous level and interface measurement in liquids.
Specs at a glance
* Process temperature:-80 °C ... +200 °C (-112 °F ... +392 °F)
* Process pressure / max. overpressure limit :Vacuum ... 100 bar (Vacuum ... 1450 psi)
* Process pressure / max. overpressure limit :Vacuum ... 100 bar (Vacuum ... 1450 psi)
Benefits
* Cost savings thanks to easy and fast commissioning as calibration takes place at the press of a button
* Reliable and safe measurement due to active build-up compensation
* Reliable and universal application thanks to wide range of certificates and approvals
* Short reaction time
* Material in contact with the process made of corrosion-resistant and FDA-listed materials
* Two-stage overvoltage protection
* No need for recalibration after replacing electronics
* Reliable and safe measurement due to active build-up compensation
* Reliable and universal application thanks to wide range of certificates and approvals
* Short reaction time
* Material in contact with the process made of corrosion-resistant and FDA-listed materials
* Two-stage overvoltage protection
* No need for recalibration after replacing electronics
Application
Liquicap FTI51 is a rod probe for point level detection of liquids that are highly viscous and tend to form build-up. Also used for detection of interface and foam of conductive liquids. Due to its active build-up compensation FTI51 offers maximum reliability and is suited for a wide range of applications thanks to international certificates. The parts in contact with the process are made of corrosion-resistant and FDA-listed materials - this guarantees a high safety standard.
For liquids that tend to form build-up. Interface detection of different liquids. Two-point control (pump control) with just one process connection. Foam detection of conductive liquids.
* Process connections: Flanges, threads, special hygienic process connections
* Temperature: -80 to +200°C (-112 to +392°F)
* Pressure: Vacuum to +100bar (vacuum to +1,450psi)
* Sensor length: 6m (20ft)
* International explosion protection certificates, overfill prevention WHG, SIL, hygienic certificates, marine approval
For liquids that tend to form build-up. Interface detection of different liquids. Two-point control (pump control) with just one process connection. Foam detection of conductive liquids.
* Process connections: Flanges, threads, special hygienic process connections
* Temperature: -80 to +200°C (-112 to +392°F)
* Pressure: Vacuum to +100bar (vacuum to +1,450psi)
* Sensor length: 6m (20ft)
* International explosion protection certificates, overfill prevention WHG, SIL, hygienic certificates, marine approval
Specification
Measuring principle
|
Capacitive Liquid
|
Characteristic / Application
|
Capacitive rod probe
|
Specialities
|
Interface detection
Foam detection |
Ambient temperature
|
-50 °C ... +70 °C
(-58 °F ... +158 °F) With separate housing: -50 °C ... +120 °C (-58 °F ... +248°F) |
Process temperature
|
-80 °C ... +200 °C
(-112 °F ... +392 °F) |
Process pressure / max. overpressure limit
|
Vacuum ... 100 bar
(Vacuum ... 1450 psi) |
Main wetted parts
|
Insulation material: PTFE, PFA
316L |
Process connection
|
G 1/2 ... 1 1/2
NPT 1/2 ... 1 1/2 Flange from DN25.../ASME 1".../JIS 10K... |
Process connection hygienic
|
Tri-Clamp ISO2852 gap free plated
Dairy coupling Flush mount |
Sensor length
|
Total length of probe: max 6 m (20 ft)
|
Communication
|
2- wire 19-253 VAC
3-wire PFM Relay DPDT DC-PNP 8/16 mA |
Certificates / Approvals
|
ATEX, FM, CSA, CSA C/US, IEC Ex, TIIS, INMETRO, NEPSI, EAC
|
Safety approvals
|
Overfill prevention WHG
SIL |
Design approvals
|
EN 10204-3.1
NACE MR0175 |
Marine approval
|
GL/ ABS/ DNV
|
Options
|
Inactive length
Fully isolated active build-up compensation Ground tube Separate housing |
Components
|
Transmitter
PFM: FTC325, FTC625, FTC470Z, FTC471Z 3-wire: FTC325 |
Application limits
|
Notice the pressure and temperature derating.
Notice for non conductive medium the DK. |
please contact email for more about the shipping cost
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.
We normally ship by DHL, UPS, FEDEX or TNT,
some country could not be ship by above method, then we would use EMS or Country Logistics line.