Overview
100% compliance, 0% effort Hygienic compact RTD thermometer, US style.
Specs at a glance
Response time
t50 = 2,5 s
t90 =9,5s
Max. process pressure (static)
at 20 °C: 40 bar (580 psi)
Operating temperature range
PT 100:
-40 °C …160 °C (-40 °F …320 °F),
optional up to 190 °C (374 °F)
Max. immersion length on request
up to 28'' (711 mm)
others on request
Response time
t50 = 2,5 s
t90 =9,5s
Max. process pressure (static)
at 20 °C: 40 bar (580 psi)
Operating temperature range
PT 100:
-40 °C …160 °C (-40 °F …320 °F),
optional up to 190 °C (374 °F)
Max. immersion length on request
up to 28'' (711 mm)
others on request
Field of application
The award-winning smart temperature probe with integrated HART® transmitter and Heartbeat Technology features the world’s first
RTD sensor unit capable of fully automated in-situ self-calibration. It effectively eliminates the risk of undetected non-conformities, reduces production downtime and increases product safety and process efficiency in applications of the food & beverage and life sciences industries. The device is fully compliant to FDA 21 CFR Part 11 regulations and GMP rules.
* Specially designed for use in hygienic and aseptic applications in the food & beverage and life sciences industries
* Measuring range: –40 to 190 °C (–40 to 374 °F)
* Pressure range up to 50 bar (725 psi)
* Protection class: IP65/67 or IP69
* Communication: Analog output 4 to 20 mA, HART® protocol
* Measuring range: –40 to 190 °C (–40 to 374 °F)
* Pressure range up to 50 bar (725 psi)
* Protection class: IP65/67 or IP69
* Communication: Analog output 4 to 20 mA, HART® protocol
Benefits
* Risk and cost reduction thanks to self-calibration and Heartbeat Technology
* No production downtime due to an automated and fully traceable inline self-calibration
* Automatized documentation, memory for 350 calibration points.
Printable calibration certificate - audit proof.
* Automated 4 to 20 mA loop check for saving time and increasing safety in your process.
* Elimination of non-conformities or undetected failures
* International certifications and approvals:
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
* Highest accuracy of measuring point through sensor-transmitter matching
* Automatized documentation, memory for 350 calibration points.
Printable calibration certificate - audit proof.
* Automated 4 to 20 mA loop check for saving time and increasing safety in your process.
* Elimination of non-conformities or undetected failures
* International certifications and approvals:
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
* Highest accuracy of measuring point through sensor-transmitter matching
Specification
Brand Name
|
ENDRESS+HAUSER
|
Model Number
|
TM372
|
Theory
|
Temperature Transmitter
|
Measuring principle
|
Resistance Temperature Detector
|
Measuring range
|
–40 to 190 °C (–40 to 374 °F)
|
Pressure range
|
at 20 °C: 40 bar (580 psi)
|
Protection class
|
IP65/67 or IP69
|
Response time
|
t50 = 2,5 s
t90 =9,5s |
Operating temperature range
|
PT 100:
-40 °C …160 °C (-40 °F …320 °F), optional up to 190 °C (374 °F) |
Characteristic Application
|
self-calibrating
imperial style compact temperature probe hygienic design hygienic process connection |
Thermowell protection tube
|
without
incl. thermowell incl. Tee-/elbow thermowell |
Insert probe
|
pipe version, isolated wires, not flexible
|
Outer diameter protection tube Insert
|
1/4'' (6,35 mm)
3/8'' (9,53 mm) 1/2" (12,7 mm) |
Max. immersion length on request
|
up to 28'' (711 mm)
|
Material protection tube/ thermowell
|
316L
|
Process connection
|
Weld-in adapter
Tri-clamp Clamp-connections according to ISO2852 G3/4" Liquiphant G1" Liquiphant Varivent T- and elbow thermowell |
Tip shape
|
straight
reduced |
Integration head transmitter
|
no (4…20mA Signal)
|
Ex - approvals
|
CSA C/US IS, I/1+2/A-F
|
International certifications and approvals:
|
– EHEDG, ASME BPE, FDA, 3-A, 1935/2004, 2023/2006, 10/2011, CE CRN, CSA General Purpose, Explosion protection, e.g. ATEX/IECEx
|
пожалуйста, свяжитесь по электронной почте для получения дополнительной информации о стоимости доставки
Обычно мы отправляем по DHL, UPS, FEDEX или TNT,
В некоторые страны невозможно отправить товар указанным выше способом, тогда мы будем использовать линию EMS или Country Logistics.
Обычно мы отправляем по DHL, UPS, FEDEX или TNT,
В некоторые страны невозможно отправить товар указанным выше способом, тогда мы будем использовать линию EMS или Country Logistics.