ROSEMOUNT 3900 General Purpose pH/ORP Sensor with Variopol (VP) Connection

: In stock

: Liquid Analyzers

: ROSEMOUNT

Precio habitual Oferta

Overview Rosemount™ 3900 General Purpose pH/ORP Sensor with Variopol (VP) Connection The Rosemount 3900VP General Purpose pH Sensor / ORP Sensor features a double junction reference that aids...

Overview

Rosemount™ 3900 General Purpose pH/ORP Sensor with Variopol (VP) Connection
The Rosemount 3900VP General Purpose pH Sensor / ORP Sensor features a double junction reference that aids in the pH/ORP sensor's resistance to harsh environments and helps prolong sensor life. The double junction protects the reference element from poisoning ions - such as ammonia, chlorine, cyanides, and sulfides. The pH/ORP sensor is made with a porous Teflon® junction. The robust pH/ORP sensor design makes the Rosemount 3900VP pH/ORP Sensor a reliable choice for a wide range of aqueous solutions found in pipelines, open tanks or ponds.

Benefits

* Extended pH sensor/ORP sensor life and protection against poisoning ions using a double junction reference
* Enhanced performance and increased life with cracking resistant glass
* Maximum chemical resistance provided by a rugged polyphenylene sulfide body
* Operates in sub-zero temperatures down to -10°C (14°F)
* Built-in solution ground for advanced diagnostics
* pH/ORP sensor body features a one-piece construction with both front facing and rear facing process threads
* ¾” and 1” MNPT process connections to meet a variety of application installation requirements
* Variopol (VP8) cable connection option, for quick cable-to-sensor release, eliminates cable twisting
* SMART enabled for pH measurements

Specification

Warranty

1 Year
Customized support

OEM, ODM, OBM
Place of Origin

United States
Brand Name

EMERSON/ROSEMOUNT
Model Number

3900 General Purpose pH/ORP Sensor with Variopol (VP) Connection
Measurement range  pH

0 to 14
Measurement range   Oxidation reduction potential (ORP)

-1500 to +1500 mV
Operating temperature

14 to 212 °F (-10 to 100 °C), automatic temperature compensation from 14 to 212 °F (-10 to 100 °C)
Maximum pressure

100 psig (790 kPa) at 212 °F (100 °C) CRN rating: 60 psig up to 212 °F (100 °C)
Materials of construction
Sensor body
Ryton® : polyphenylene sulfide (PPS)

O-ring
EPDM

pH electrode
Glass

ORP electrode
Glass, platinum

Solution ground
Stainless steel

Reference junction
PTFE
Conductivity

Responds to changes in pH at a minimum conductivity of 0.1 µS/cm when used with the low flow panel. The sample flow rate must be
controlled to 2 gallons per hour (7.61 liters per hour).
Process connection
Front facing
¾-in. and 1-in. male national pipe thread (MNPT)

Rear facing
1-in. MNPT
Temperature compensation

Pt-100