YOKOGAWA Multi Channel 4-Wire Analyzer FLXA402

: In stock

: Liquid Analyzers

: YOKOGAWA

Precio habitual Oferta

Overview Multi Channel 4-Wire Analyzer FLXA402 The FLXA402 offers multiple connection possibilities, eliminating the need for multiple analyzers and providing greater flexibility. This four-wire analyzer includes a color HMI...

Overview

Multi Channel 4-Wire Analyzer FLXA402
The FLXA402 offers multiple connection possibilities, eliminating the need for multiple analyzers and providing greater flexibility. This four-wire analyzer includes a color HMI and an easy touchscreen operation with a simple menu structure in 9 languages for simple and efficient setup and configuration.
With its modular design, the FLXA402 analyzer offers a range of measurement choices with the respective sensor module, including pH/ORP, resistivity/conductivity (SC), inductive conductivity (ISC), percent concentration, dissolved oxygen (DO), and 4-20mA
input. Dual sensor measurements offer additional functionalities, such as a calculated data function that can be customized.

Specification

Measurement Object
• pH/Oxidation-reduction Potential (pH/ORP) • Conductivity (SC) • Inductive Conductivity (ISC) • Dissolved Oxygen (DO) Note: The
available measurement object depends on a sensor module installed on the converter.
Types of Sensor Module
Analog sensor module for analog sensors • PH: for analog pH/ORP • SC: for analog Resistivity/ Conductivity • ISC: for analog
Inductive Conductivity • DO: for analog Dissolved Oxygen Digital sensor module • SENCOM SA: for SA11 • Digital sensor: for optical
DO
Other Modules
• IO: mA output, mA input, contact input • Relay: SPDT relay • Digital communication: Modbus TCP/IP(Ethernet) or Modbus RTU
(RS485)
Combination of Sensor Module
There are two sensor module slots; up to Five sensor measurements are available in case of the combination of SENCOM SA module and
BA11 junction box. When FLXA402 (-DD) is selected, multiple sensor measurements are not available. The combination of two same
sensor modules or different type of sensor modules is possible except in the case of Optical DO sensor.
 Input Specification
Dual high impedance input (≥1012 Ω)
■ Input Range
pH: -2 to 16 pH (with option /K: 0 to 14 pH) ORP: -1500 to 1500 mV rH: 0 to 100 rH Temperature: Pt1000: -30 to 140 ºC Pt100: -30
to 140 ºC 6k8: -30 to 140 ºC PTC10k: -30 to 140 ºC NTC 8k55: -10 to 120 ºC 3k Balco: -30 to 140 ºC PTC500: -30 to 140 ºC
Output Range
■ Output Range pH: min. span 1 pH max. span 20 pH ORP: min. span 100 mV max. span 3000 mV rH: min. span 2 rH max. span 100 rH
Temperature: min. span 25 ºC max. span 170 ºC
Performance (Accuracy)
(The specifications are expressed with simulated inputs.) pH Linearity: ±0.01 pH Repeatability: ±0.01 pH Accuracy ±0.01 pH ORP
Linearity: ±1 mV Repeatability: ±1 mV Accuracy: ±1 mV Temperature with Pt1000, 6k8, PTC10k, NTC 8k55, 3k Balco, PTC500 Linearity:
±0.3 ºC Repeatability: ±0.1 ºC Accuracy: ±0.3 ºC with Pt100 Linearity: ±0.4 ºC Repeatability: ±0.1 ºC Accuracy: ±0.4 ºC