Overview
Compact pressure sensor with the Contite measuring cell for hydrostatic level measurement.
Specs at a glance
* Accuracy :Standard 0.2% Optional 0.1%
* Process temperature :-10°C...100°C (14°F...212°F)
* Pressure measuring range :100 mbar...10 bar (1.5 psi...150 psi)
* Process pressure / max. overpressure limit : 40 bar (600 psi)
* Main wetted parts : Alloy C 316L optional coating AuPt
* Max. measurement distance :100m (328ft) H2O
* Material process membrane : 316L, AlloyC Gold-Rhodium
* Measuring cell : 100 mbar...10 bar (1.5 psi...150 psi)
* Process temperature :-10°C...100°C (14°F...212°F)
* Pressure measuring range :100 mbar...10 bar (1.5 psi...150 psi)
* Process pressure / max. overpressure limit : 40 bar (600 psi)
* Main wetted parts : Alloy C 316L optional coating AuPt
* Max. measurement distance :100m (328ft) H2O
* Material process membrane : 316L, AlloyC Gold-Rhodium
* Measuring cell : 100 mbar...10 bar (1.5 psi...150 psi)
Benefits
* Hermetically sealed Contite measuring cell with condensate-resistance, high reference accuracy: ±0.2%, optionally ±0.1% and minimum temperature effects
* Modular concept for easy replacement of display or electronics
* Seamless and independent system integration (HART/PA/FF)
* Easy menu-guided commissioning via local display, 4 to 20mA with HART, PROFIBUS PA, FOUNDATION Fieldbus
* International usage thanks to a wide range of approvals
* Modular concept for easy replacement of display or electronics
* Seamless and independent system integration (HART/PA/FF)
* Easy menu-guided commissioning via local display, 4 to 20mA with HART, PROFIBUS PA, FOUNDATION Fieldbus
* International usage thanks to a wide range of approvals
Application
The compact Deltapilot FMB50 transmitter with the condensate resistant Contite measuring cell is typically used in the process and hygienic applications.It is made for level measurement in liquid and paste-like media in open or closed containers and unaffected by possible foam formation. For use in SIL2 safety systems.For level and/or pressure measurement.
* Process connections: Thread, flanges and flush-mounted hygienic connections
* Process temperature: –10 to +100°C (14 to 212°F) (+135°C (275°F) for 30 minutes maximum)
* Measuring range: 0.1 to 10bar (1.5 to 150psi)
* Accuracy: ±0.2%, ±0.1% (option)
* International explosion protection certificates, overfill prevention WHG, SIL, hygienic certificates
* Process connections: Thread, flanges and flush-mounted hygienic connections
* Process temperature: –10 to +100°C (14 to 212°F) (+135°C (275°F) for 30 minutes maximum)
* Measuring range: 0.1 to 10bar (1.5 to 150psi)
* Accuracy: ±0.2%, ±0.1% (option)
* International explosion protection certificates, overfill prevention WHG, SIL, hygienic certificates
Specification
Measuring principle
|
Hydrostatic pressure
|
Characteristic
|
Compact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences |
Supply voltage
|
4...20 mA HART
10,5...45 VDC (Non Ex): Ex ia: 10,5...30 VDC PROFIBUS PA: 9...32 VDC (Non Ex) FOUNDATION Fieldbus: 9...32 VDC (Non Ex) |
Reference Accuracy
|
Standard 0.2 %
Optional 0.1 % |
Long term stability
|
0.05 % of URL/ year
|
Process temperature
|
-10°C...100°C
(14°F...212°F) |
Ambient temperature
|
-40°C...85°C
(-40°F...185°F) |
Measuring cell
|
100 mbar...10 bar
(1.5 psi...150 psi) |
Smallest calibratable span
|
10 mbar (1.45 psi)
|
Vacuum resistance
|
0 mbar abs.
|
Max. Turn down
|
100:1
|
Max. overpressure limit
|
40 bar (600 psi)
|
Process connection
|
Thread:
G1 1/2, MNPT1 1/2 Flange: DN40...DN100, ASME 1 1/2"...4", JIS 10K |
Process connection hygienic
|
Tri-Clamp
DIN11851 DIN11864-1 NEUMO Varivent N SMS DRD |
Material process membrane
|
316L, AlloyC
Gold-Rhodium |
Material gasket
|
None, measuring cell welded
|
Fill fluid
|
Inert oil,
Synthetic oil |
Material housing
|
316L, Die-cast aluminum
|
Communication
|
4...20 mA HART
PROFIBUS PA FOUNDATION Fieldbus IO-Link |
Certificates / Approvals
|
ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
|
Safety approvals
|
Overfill protection WHG
SIL |
Design approvals
|
EN10204-3.1
NACE MR0103 |
Hygienic approvals
|
3A, EHEDG
CoC ASME-BPE |
Marine approvals
|
GL/ ABS/ LR/ BV/ DNV
|
Drinking water approvals
|
KTW/ NSF/ ACS
|
Specialities
|
Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions |